Interference of hemoglobin variants with HbA1c measurements: comparison of 6 commonly used HbA1c methods with the IFCC reference method
Mingyang Li, Song Ge, Xin Shu, Xiongjun Wu, Haiyan Liu, Anping Xu, Ling JiAbstract
Background
Glycated hemoglobin, or hemoglobin A1c (HbA1c), serves as a crucial marker for diagnosing diabetes and monitoring its progression. We aimed to assess the interference posed by common Hb variants on popular HbA1c measurement systems.
Methods
A total of 63 variant and nonvariant samples with target values assigned by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method were included. We assessed 6 methods for measuring HbA1c in the presence of HbS, HbC, HbD, HbE, and fetal hemoglobin (HbF): 2 cation-exchange high-performance liquid chromatography (HPLC) methods (Bio-Rad D-100 and HLC-723 G8), a capillary electrophoresis (CE) method (Sebia Capillarys 3 TERA), an immunoassay (Roche c501), an enzyme assay system (Mindray BS-600M), and a boronate affinity method (Primus Premier Hb9210).
Results
The HbA1c results for nonvariant samples from the 6 methods were in good agreement with the IFCC reference method results. The Bio-Rad D-100, Capillarys 3, Mindray BS-600M, Premier Hb9210, and Roche c501 showed no interference from HbS, HbC, HbD, and HbE. Clinically significant interference was observed for the HLC-723 G8 standard mode. Elevated HbF levels caused significant negative biases for all 6 methods, which increased with increasing HbF concentration.
Conclusion
Elevated levels of HbF can severely affect HbA1c measurements by borate affinity, immunoassays, and enzyme assays.