Comparative sensitivity and accuracy of 24 vs. 18 MHz ultrasound probes for the diagnosis of flexor tendon repair integrity

Sean Renfree, Nathaniel B. Hinckley, Nirvikar Dahiya, Nan Zhang, Kevin J. Renfree
  • Radiology, Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Abstract

Purpose

To determine whether a 24 MHz transducer significantly improves sensitivity, specificity and accuracy in evaluating flexor tendon repair integrity compared with an 18 MHz transducer.

Methods

One hundred and twelve cadaveric digits were randomised to an intact repair or simulated ‘failed’ repair, and to a two‐ or eight‐strand repair of a flexor digitorum profundus laceration. A blinded sonologist evaluated specimens in static mode using 18 and 24 MHz transducers. Gaps were remeasured after scanning, and final gap width recorded. McNemar's exact test calculated differences between sensitivity, specificity and accuracy, and chi‐squared test to compare sensitivity, specificity and accuracy between number of strands (2 vs. 8) and repair gap (≥4 mm).

Results

The 24 MHz transducer had higher sensitivity (81 vs. 59%), lower specificity (67 vs. 70%) and higher overall accuracy (74 vs. 64%), than the 18 MHz transducer. The difference for sensitivity was significant (P = 0.011), but not differences for specificity and overall accuracy (P > 0.05). Pearson's correlation (r = 0.61) demonstrated a moderate‐to‐strong positive correlation between measured and true gap sizes. Increased number of suture strands (2 vs. 8) did not impair sensitivity, specificity nor accuracy.

Discussion

Ultrasound may tend to overestimate gap width, and a slight risk that some intact repairs, or those with small, clinically insignificant gaps may undergo surgical exploration that may not be indicated.

Conclusions

A 24 MHz transducer is a more sensitive and accurate transducer for assessing flexor tendon repair integrity and measuring small gaps.