DOI: 10.1371/journal.pone.0353640 ISSN: 1932-6203

Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter

Devesh Kumar, Sunita Kapoor, Ambanna Gowda, Dipti Gupta, Harsh Mittal, Suneet Sood

Background

Non-invasive hemoglobin measurement offers a painless and rapid alternative to conventional blood-based testing. The Vivaray Hb pro is a handheld photoplethysmography-based device designed for point-of-care hemoglobin assessment without blood sampling. We evaluated the clinical performance of the Vivaray Hb pro by comparing device-generated hemoglobin values with those obtained from a calibrated laboratory blood cell counter.

Methods

In this cross-sectional, non-randomized clinical performance study, participants aged ≥8 years were prospectively recruited. Hemoglobin was measured non-invasively using the Vivaray Hb pro and compared with venous blood samples analyzed on a calibrated Coulter counter. Agreement between methods was assessed using Bland–Altman analysis, including regression-based evaluation for proportional bias. Mean difference, mean absolute error (MAE) and proportions of measurements within tolerance limits were also calculated. Complete paired measurements were available for 763 individuals.

Results

Bland–Altman analysis returned a mean difference of −0.40 g/dL. The bias was hemoglobin-dependent, being higher than the true value at low hemoglobin levels and underestimation at higher levels. Regression-based analysis showed proportional bias (β 1  = −0.178), indicating decreasing difference with increasing hemoglobin concentration. The MAE was 1.5 g/dL but was lower (1.2) in the clinically predominant ranges of 8.1–13 g/dL.

Conclusion

The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.