Use of a novel tapered, scoring balloon to treat juxta-anastomotic stenosis in AV fistula

Xiangjie Li, Zhe Wang, Xia Zhao, Hao Sun, Qining Fu
  • Nephrology
  • Surgery

Background:

Juxta-anastomotic stenosis, the most common lesion in arteriovenous fistula, creates a dilemma for optimal balloon size choosing during PTA.

Objectives:

To descript the effect of a novel tapered scoring balloon catheter (DK Medtech, Suzhou, China) which has a conical shape with an increasing diameter from the front part of the balloon to the end, and three non-slip elements attached on the surface of the balloon.

Research design:

Case series of 10 patients used this balloon catheter was retrospectively analyzed.

Subjects:

Patients with juxta-anastomotic stenosis.

Measures:

A retrospective review of 10 cases using the novel tapered, scoring balloon catheter in our center was performed.

Results:

All cases were clinical technique success. The average total procedure time was 16.7 min with 2:14 of fluoroscopy time. No complications such as vascular rupture or dissection occurred. The primary patency rates at 6 and 12 month were 80% and 50% separately.

Conclusions:

This tapered scoring balloon provides an economical, safe, and efficient management for juxta-anastomotic stenosis.