DOI: 10.3390/jcs10110588 ISSN: 2504-477X

Tunable Laser-Reduced Graphene Oxide–DNA Aptamer Composite for Rapid Disease Detection

Olga M. Antipova, Mariya I. Trukhanova, Egor A. Danilov, Aleksei V. Shchegolkov, Nina V. Myasishcheva, Ivan I. Bobrinetskiy, Ivan A. Komarov

Flexible graphene-derived composites are promising transduction materials for rapid biosensing, but conductivity and surface reactivity must be balanced within the same device. Here, a laser-patterned reduced graphene oxide (rGO)/poly(ethylene terephthalate) (PET) composite was developed for label-free resistive detection of herpes simplex virus type 1 (HSV-1) glycoprotein D (gD). A graphene oxide film was spin-coated on PET and locally reduced at 48.3 mJ cm−2 to form conductive contacts and at 25 mJ cm−2 to create a sensing region retaining carboxyl groups. A 5′-amine-modified 44-nucleotide DNA aptamer was covalently immobilized by carbodiimide coupling. Atomic force microscopy indicated a local film thickness of approximately 62–72 nm, while infrared and Raman spectroscopy supported the intended chemical reduction gradient. Aptamer immobilization increased sensor resistance by approximately 5–7%. Across 15 sensors, gD produced a concentration-dependent response that was clearly separated from albumin; a representative endpoint response was approximately 6.5% for gD and 0.7% for albumin. The lowest demonstrated detection level was approximately 100 pM, without a formal statistical LOD calculation. The approximate semilog-linear interval was 1 nM–1 mM, and the response time was 70–100 s. These results demonstrate that spatially programmed laser reduction can integrate conductive and bioreactive functions in a composite-based biosensor on a flexible substrate.

More from our Archive