Torque deviation at different angulations of dynamic abutments with narrow implants in the esthetic zone
Mohamed F. Kandil, Nahed A.K. Kashef, Amal M. BasyouniBackground
Screw-retained restorations in the esthetic zone may present biomechanical challenges when dynamic abutments with different angulations are used, as torque deviation and screw loosening can affect clinical performance.
Purpose
To evaluate tightening torque deviation and screw loosening for different angles of dynamic abutments used with narrow implants in the esthetic zone.
Patients and methods
Twenty epoxy resin blocks were assigned to two groups (
Results
Both groups demonstrated removal torque loss before and after cyclic loading. No significant difference was found between 10 and 25° abutments regarding screw loosening (
Conclusion
Within the limitations of this in-vitro study, both dynamic abutment angulations demonstrated similar removal torque loss with no increased risk of screw loosening. Significant torque deviation was associated with the hexalobular screw system.