Swinging vs Flicking: Assessing Safer Removal Techniques for Air Bubbles in Intravitreal Injections
Paula Sakemi Fukuhara, Alexandra Lucas, Krishna M.G. Mallela, Øystein Kalsnes Jørstad, Michel Eid Farah, Gustavo Barreto MeloPurpose:
To evaluate whether a gentler swinging motion, rather than flicking, can effectively eliminate air bubbles from syringes before intravitreal injections while minimizing particle generation.
Methods:
Three syringe models were assessed under controlled conditions: Zero Residual Silicone Oil (SO)-free 0.2 mL syringe, BD Tuberculin 1 mL slip-tip syringe (siliconized syringe 1), and HSW Soft-Ject 1 mL Luer-Lock syringe (siliconized syringe 2). Two solutions were tested: a polysorbate-based buffer and bevacizumab (Avastin). Each syringe was subjected to 3 handling conditions: no agitation (control), swinging (hand-induced centrifugal movement), and flicking. The number of released particles was quantified using FlowCam imaging, and statistical analyses were conducted to compare handling techniques and syringe models.
Results:
Flicking significantly increased particle release compared with both swinging and the control condition (
Conclusions:
Compared with flicking, a gentle swinging motion reduced particle release across syringe models while allowing air bubbles to be removed. This technique may represent a simple modification to intravitreal injection preparation that could reduce particulate contamination.