Single-Step Microfluidic Anti-Factor Xa Assay: Development and Clinical Evaluation in Patients on Low-Molecular-Weight Heparin
Stacia Shi En Lim, Yuxi Wang, Jasmine Shu Hui Lim, Jeanie Jing Yi Yeo, Christina Lai Lin Sum, Mui Ka Ang, Bernard Pui Lam Leung, Bingwen Eugene Fan, Yi FuDirect oral anticoagulants and low-molecular-weight heparins (LMWHs) act principally through inhibition of activated factor X (FXa), yet no point-of-care assay for anti-FXa activity is in routine clinical use. This study reports a disposable microfluidic device that performs a single-step colorimetric anti-FXa assay on plasma. The chip was fabricated by laminating laser-cut poly(methyl methacrylate) sheets with adhesive tape; both assay reagents were pre-stored on-chip in dried form, and absorbance was measured at 405 nm. In 58 plasma samples from patients receiving LMWH therapy, the on-chip result correlated moderately with the automated laboratory assay. Benchmarking against the same reagents in a microwell plate showed that the device recovered 46% of the attainable agreement. The microfluidic assay identified samples at or above a therapeutic anti-Xa concentration with 94% specificity and a positive predictive value of 96%. The device is therefore suited to rapid screening prior to laboratory confirmation.