Radiation in the Bronchoscopy Suite
Wilson S. Tsai, Erin Haywood, Xinhua Li, Jeremy Rosenbaum, Brenna LindseyBackground:
This study aimed to quantify radiation doses during navigational bronchoscopy procedures, comparing them with reported cohorts and evaluating the LungVision (Body Vision Medical Inc.) system’s efficacy in dose reduction.
Methods:
This retrospective observational study included 52 consecutive navigational bronchoscopy cases, categorized into 4 imaging groups based on the C-arm: Cios Spin (Siemens Healthineers), or OEC 9900 (GE HealthCare); and the 3D tomographic imaging algorithm: Cios Spin’s onboard imaging, or LungVision’s AI-driven imaging. Patient and lesion data, outcomes, and radiation indices were collected. Existing literature on 3D image guidance for bronchoscopic lung nodules was reviewed to compare reported radiation doses.
Results:
Combining LungVision with Cios Spin significantly reduced radiation dose in all cases compared with using Cios Spin alone: Cumulative air kerma (Ka,r) reduced from 238.7 to 119.1 mGy (
Conclusion:
Following the “as low as reasonably achievable” (ALARA) principle minimizes ionizing radiation exposure, benefiting patients and operators. Physicians should compare baseline radiation levels with the literature and adopt dose-reduction techniques. LungVision’s lower dose indices render it effective for real-time 3D imaging during navigational bronchoscopy while reducing radiation dose.