Optimizing oxygen delivery in duckbill bag-valve-mask devices during spontaneous breathing
Wai-Hung Yu, Kwok Ming Ho, Yu-Yeung YipIntroduction
Pre-oxygenation is often performed with bag-valve-mask (BVM) devices whose oxygen delivery performance during spontaneous breathing is incompletely characterized. Duckbill-valve BVMs without an expiratory valve are prone to air entrainment during inspiration. This study evaluated how patient valve configuration and ventilatory pattern influence oxygen delivery during simulated spontaneous breathing with two duckbill devices.
Methods
In a mechanical lung simulator, we tested two adult BVMs: a duckbill device without an expiratory valve (Med-Rescuer) and one with a disc-type expiratory valve (Laerdal The BAG II). Devices were supplied with oxygen flows of 8, 15, or 30 L/min. Spontaneous breathing patterns were programmed across a broad range of volumes (250–1000 mL) and rates (15 and 25 breaths per minute). Additional runs for Med-Rescuer included a PEEP valve (5 and 10 cm H 2 O) to reduce air entrainment.
Results
For Med-Rescuer, delivered oxygen fraction (
Conclusions
The duckbill BVM incorporating an expiratory valve delivered higher and more stable