Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements

Jie Hou, Sisay Mebre Abie, Ørjan G. Martinsen

Abstract

Phase angles measured at low frequencies with the tetrapolar technique are sometimes scattered or even positive. We used Monte Carlo simulations of lock-in detection to quantify how large phase errors that measurement noise alone can cause. For a signal-to-noise ratio after demodulation (SNR out ) of about 3 or higher, the standard deviation of the phase error was close to 57° divided by SNR out , and 95% of the errors were within about twice this value. At SNR out = 10 the error standard deviation was 6°; at SNR out = 1 it was 64° and 16% of the measurements were wrong by more than 90°. As SNR out decreased towards zero, the phase gradually approached a completely random value. In an illustrative low-frequency tetrapolar measurement, the absolute phase errors were small (standard deviation 0.007° at 1 Hz and 0.04° at 0.1 Hz), but the true tissue phase was even smaller. The error was therefore 1.2 times the true phase at 1 Hz and 37 times at 0.1 Hz, and the measured phase had the wrong sign in 20% and 49% of the measurements, respectively. The sign of a small phase angle is only reliable (wrong in less than 1% of measurements) when the true phase is larger than 2.3 times the error standard deviation.