In Vitro Wear Simulation of a Kinematically Aligned Knee Implant Tilted at an Angle of 8°

Stefan Schroeder, Mareike Mueller, Maximilian Uhler, Therese Bormann, Sebastian Jaeger, Jan Philippe Kretzer

ABSTRACT

Despite promising results of kinematic alignment (KA), there are still concerns regarding the wear behavior of KA knee implants. In a previous study, mechanically aligned (MA), KA and malaligned (MalA) knee implants were tested in a wear simulation study using Attune knee implant systems. The results showed no increased wear rate for a mild KA (KA4°) and MalA (MalA4°) condition in comparison to MA (MA0°) knee implants. This study analyzes the wear behavior of a KA condition of 8° (KA8°). The test setup was adapted accordingly. In addition, all tests of the previous study (MA0°, MalA4° and KA4°) and the KA8° condition were replicated using PFC implant systems to indicate if the previous results can be transferred to the predecessor model of the Attune knee implant system. The KA8° using the Attune knee implant system did not lead to a higher wear rate (2.4 ± 0.1 mg/10 6 cycles) in comparison to the wear results of the previous study (3.8 ± 0.5 mg/10 6 cycles for MA0°; 2.7 ± 0.2 mg/10 6 cycles for MalA4° and 4.1 ± 0.2 mg/10 6 cycles for KA4°). The PFC implant system led to higher wear rates than the Attune implant system but showed the same trend for the different alignment methods (5.1 ± 0.4 mg/10 6 cycles for MA0°; 3.9 ± 0.1 mg/10 6 cycles for MalA4°, 5.1 ± 0.8 mg/10 6 cycles for KA4° and 3.5 ± 1.0 mg/10 6 cycles for KA8°). KA8° did not show wear related problems but further in vitro and in vivo studies are necessary.