In vitro bench testing using patient-specific 3D models for percutaneous pulmonary valve implantation with Venus P-valve
Yu Han, Zehua Shao, Zirui Sun, Yan Han, Hongdang Xu, Shubo Song, Xiangbin Pan, Peter P. T. de Jaegere, Taibing Fan, Gejun Zhang- General Medicine
- General Medicine
Abstract
Background:
Due to the wide variety of morphology, size, and dynamics, selecting an optimal valve size and location poses great difficulty in percutaneous pulmonary valve implantation (PPVI). This study aimed to report our experience with
Methods:
Patient-specific 3D soft models were generated using PolyJet printing with a compliant synthetic material in 15 patients scheduled to undergo PPVI between July 2018 and July 2020 in Central China Fuwai Hospital of Zhengzhou University.
Results:
3D model bench testing altered treatment strategy in all patients (100%). One patient was referred for surgery because testing revealed that even the largest Venus P-valve would not anchor properly. In the remaining 14 patients, valve size and/or implantation location was altered to avoid valve migration and/or compression coronary artery. In four patients, it was decided to change the point anchoring because of inverted cone-shaped right ventricular outflow tract (RVOT) (