Impact of aortic annulus size on clinical and hemodynamic outcomes of transcatheter aortic valve implantation with the navitor valve

J Capparelli, M Casenghi, S Corradetti, P Paolisso, M Belmonte, E Gallinoro, E Fiori, L Piras, P Policastro, C Barsali, F Cefali', G Vincelli, D Ausiello, L Rosseel, E Barbato

Abstract

Background

Patients with small aortic annuli undergoing transcatheter aortic valve implantation (TAVI) are at risk of higher residual transvalvular gradients. Evidence on the performance of new-generation intra-annular self-expanding valves in this anatomical setting remains limited. We evaluated the impact of annular size on procedural, hemodynamic, and clinical outcomes after TAVI with the Navitor valve.

Methods

Patients were stratified by annular area measured by multi-slice computed tomography, defining small annulus as ≤430 mm². Primary endpoints at 12 months were: (i) a clinical composite of all-cause death, disabling stroke, or heart failure rehospitalization and (ii) a composite endpoint of bioprosthetic valve dysfunction. Secondary endpoints included VARC-3–defined procedural and early outcomes and valve hemodynamic performance.

Results

The study population consisted in 218 patients, including 91 (41.7%) with a small annulus. VARC-3 technical success was high and comparable between small and non-small annuli (96.7% vs. 96.1%). Patients with small annuli had higher early post-procedural mean transvalvular gradients (7.7 ± 3.8 vs. 6.6 ± 3.0 mmHg; p=0.029) and smaller effective orifice areas index (1.1 ± 0.3 vs. 1.3 ± 0.2 cm²/m2; p=0.037), while the incidence of mean gradients ≥20 mmHg remained low (1.1% vs. 0.0%; p=0.432). At 12 months, the clinical composite endpoint occurred in 8.8% of patients with small annuli and 10.2% with non-small annuli (p=0.759). Bioprosthetic valve dysfunction was infrequent and similar between groups (6.6% vs. 6.3%; p=0.875).

Conclusions

TAVI with the Navitor valve was associated with high procedural success, favourable valve hemodynamics, and comparable mid-term clinical outcomes irrespective of annular size.12-months clinical and Haemodynamic Outcomes