Human factors testing in a novel automated peritoneal dialysis system
Nupur Gupta, Vikram Aggarwal, James A Sloand, Rick Pendergraft, Steve J Lindo, Brent MillerBackground
Because automated peritoneal dialysis (APD) systems are used in the home and operated by patients or their care partners, the human–device interface must be thoughtfully designed to facilitate intuitive operation to reduce the risk of errors that could occur during their use. User-friendly design should also help reduce anxiety to peritoneal dialysis (PD) adoption by patients incident to end-stage kidney disease (ESKD). Usability issues remain a significant barrier to PD adoption and are an important contributor to premature death, serious injuries, and PD technique failure.
Methods
A summative human factors usability study was conducted on a novel, gravity-based APD device (
Results
Of the Simulated Use tasks evaluated, 97.3% were deemed successful across all users. This high success rate demonstrates the effectiveness of the
Conclusion
In this summative human factors usability study, the