Historical control trial: safe tibia varus cut in kinematic total knee arthroplasty using mechanical instrument combined 3D-printed tibia recut guide
Nguyen Phuc Thinh, Pham Thi Cam Van, Bui Hong Thien KhanhBackground:
Kinematic alignment total knee arthroplasty (KA TKA) has the potential to restore native knee kinematics and improve functional outcomes compared with conventional mechanical alignment (MA) techniques; however, controlling tibial varus cuts remains a challenge, particularly in resource-limited settings.
Methods:
This study aimed to evaluate a modified KA TKA technique using conventional mechanical instruments combined with a 3D-printed tibial recut guide to control tibial varus alignment, and to compare its radiologic measures with those of MA TKA performed with navigation assistance. A historical control trial was conducted involving 100 patients. Fifty patients underwent KA TKA using mechanical instruments and a 3D-printed tibial recut guide (2020–2022), and 50 historical controls underwent MA TKA with navigation (2016–2019). Radiologic assessments included joint-line obliquity, femoral and tibial mechanical angles (TMAs), and the hip-knee-ankle (HKA) angle, measured using EOS imaging pre- and postoperatively.
Results:
Postoperative femoral mechanical angles showed no significant difference between groups (KA: 89.6° ± 2.9, MA: 89.3° ± 2.4;
Conclusions:
The modified KA TKA technique, using conventional instruments combined with a 3D-printed tibial recut guide, achieves reliable joint line restoration and controlled tibial varus cuts comparable to navigation-assisted MA TKA.