From Earlier to Updated Quadrifocal Intraocular Lens Designs: An Objective Optical Metrology Analysis
Eun Ah Jeong, Seung Pil Bang, Seth Pantanelli, Su Jin Kang, Sonia H. YooPurpose:
To compare the optical performance of the Clareon PanOptix and Clareon PanOptix Pro intraocular lenses (IOLs) using interferometric metrology and identify measurable optical differences.
Setting:
Physiological Optics Laboratory, Daegu, Republic of Korea.
Design:
Experimental optical-bench investigation.
Methods:
Three +20.0 D samples each of the PanOptix and PanOptix Pro IOLs (Alcon) were analyzed using a Mach–Zehnder interferometer in a model eye incorporating +0.27 µm corneal spherical aberration. Microscopic imaging, wavefront reconstruction with Zernike analysis, diffractive-step profile reconstruction, point-spread function (PSF)-based retinal image simulation, halo analysis, and through-focus modulation transfer function (TF-MTF) analysis were performed.
Results:
Both IOLs showed broadly similar diffractive architecture and spherical aberration profiles. Peak-to-valley step-profile amplitudes were comparable, although the PanOptix Pro showed localized differences in the reconstructed step-height contour. PSF-based analyses showed broadly similar focal behavior, but the PanOptix Pro demonstrated clearer Sloan
Conclusions:
The PanOptix Pro preserves the fundamental design and spherical aberration profile of the original PanOptix, while showing localized diffractive step-height modifications. Together with clearer far-intermediate F-image preservation, sustained central PSF-core brightness, and a shallower post-peak aMTF dip, these findings are consistent with a relative redistribution of optical energy toward the far-intermediate region, supporting greater far-intermediate optical continuity without substantial changes in nominal add power or overall focal structure.