Evaluation and comparison of the scanning accuracy of three intraoral scanners for single crown, three-unit, and cross-arch fixed partial denture by three-dimensional digital superimposition technique and measurement of vertical marginal fit of three
Mansi Dhiran, Ajay Sabane, Abhijit Ashok Patil, Arti Gachake, Darshana Mundhe, Rupali PatilAbstract
Aim:
This study aims to evaluate and compare the scanning accuracy of three IOSs and measuring the vertical marginal fit of three-dimensional (3D) printed resin retainers for single crown, three-unit, and cross-arch fixed partial denture situations.
Settings and Design:
Comparative
Materials and Methods:
A maxillary Nissin typodont with prepared abutments at #13, #23, #25, #27, and #16 served as the master model. A calibrated Shining 3D laboratory scanner provided the reference standard tessellation language (control). Each IOS generated six scans (
Statistical Analysis Used:
One-way ANOVA with Tukey’s post-hoc test was used for comparisons among scanners and prosthesis spans. Pearson’s correlation was used to assess the relationship between trueness and marginal fit, and Chi-square test was used to assess clinical acceptability. A
Results:
3Shape TRIOS 3 showed the highest trueness (28.5 ± 5.2 µm) and precision (18.2 ± 3.1 µm), followed by Medit i700 and Shining 3D Aoralscan 3 (
Conclusion:
All three scanners delivered clinically acceptable results; 3Shape TRIOS 3 yielded the highest accuracy and best marginal adaptation. Accuracy and marginal fit deteriorated as prosthesis span increased.