Design and Validation of a 3D-Printed Low-Frequency Mechanical Stimulation Device for In Vitro Applications

Veronika Sedláková, Peter Čurma, Zuzana Naďová, Norbert Ferenčík, Marianna Ondrášová, Radovan Hudák

This study describes the development, technical validation, and preliminary biological evaluation of a custom-built, low-cost device designed to apply low-amplitude mechanical vibrations to cultured cells. The system consists of 3D-printed PLA components and a miniature loudspeaker (8 Ω, 0.5 W) driven by a pulse-width modulation (PWM) module. The electrical excitation signal was verified using digital oscilloscopy and Fast Fourier Transform (FFT) analysis, which confirmed a stable fundamental frequency of 30 Hz within the mechanically effective operating range of the actuator (approximately 20–90 Hz); higher harmonic components at integer multiples of the fundamental frequency (60–180 Hz), characteristic of rectangular PWM excitation, were also detected. At the stimulation setting (30 Hz, 30% duty cycle), a MEMS accelerometer at the culture position recorded a resultant acceleration of 1.04–1.10 g (vibration-related component of approximately 0.04–0.10 g). Human gingival fibroblasts (GF06) cultured in 35 mm Petri dishes were exposed to daily vibration outside a CO2 incubator, and their response was evaluated using the MTT assay and an in vitro wound healing (scratch) assay. After 24 h of daily 15 min stimulation, the metabolic activity of vibration-treated cultures exceeded that of the controls by approximately 15 and 25 percentage points at the lower and higher seeding densities, respectively; however, these differences were not statistically significant (p = 0.189 and p = 0.098). When the daily stimulation duration was varied, the MTT signal after 72 h increased with the stimulation duration, reaching approximately 115%, 134%, and 154% of the control for 150, 300, and 450 s, respectively (p < 0.05 for 300 and 450 s). In the scratch assay, stimulation for 450 s increased gap closure compared with the control after 6 h (35.4% vs. 3.4%) and 24 h (53.4% vs. 21.2%; p < 0.01). No evident cytotoxic effects were observed under the tested conditions. The study demonstrates the technical feasibility of the platform and its ability to elicit a measurable, stimulation duration-dependent biological response, and it provides a basis for future experiments that include direct mechanical characterization of the stimulus and larger numbers of independent biological replicates under controlled physiological conditions.

More from our Archive