Coupled fatigue–wear degradation and nitinol SMA integration in a tendon-driven hand prosthesis: A predictive analytical framework for hybrid actuation
Jair Leopoldo Loaiza-Bernal, Nelson Arzola
Tendon-driven prosthetic hands fail in service because the steel cable fatigues at the guide bends while, simultaneously, the polymer channel wears at the same contact. The two mechanisms are coupled: the wear groove changes the geometry that sets the stress concentration, so fatigue accelerates on a surface that is itself degrading. Static finite-element analysis misses this interaction entirely. We develop here a purely analytical framework—Goodman fatigue criterion, Archard wear law, Monte Carlo uncertainty propagation—for a cable-driven prosthesis with Nylon 6/10 phalanges and DC gear-motor actuation; no experimental results are reported, and every prediction is formulated as a falsifiable hypothesis for subsequent testing. The model predicts a median cable life of 2.1×10
6
cycles (95% CI: 1.3–3.4×10
6
), roughly 210 days of intensive use, with 0.53 mm of cumulative channel wear. Parametric sweeps show that the guide radius and the wear coefficient