DOI: 10.1097/j.jcrs.0000000000002079 ISSN: 0886-3350

Comparison of the Moussa Intraocular Lens Formula to contemporary formulas in a multicenter study

Peter Kiraly, Hisham Hamze, Soon Ch’ng, George Moussa

Purpose:

To evaluate the refractive predictive accuracy of the Moussa intraocular lens (IOL) formula compared with Kane, EVO, PEARL-DGS, K6, Hoffer-QST, and Barrett Universal II (BUII).

Setting:

Birmingham Midland Eye Centre (Birmingham), Manchester Royal Eye Hospital and Newmedica (Manchester), UK .

Design:

Multicenter, retrospective, consecutive, nonrandomized observational case series (September 2022–June 2025).

Methods:

Included were one eye per patient undergoing uncomplicated phacoemulsification. Preoperative biometry used the IOLMaster 700, Lenstar LS900, or Pentacam AXL Wave. Lens-constants were optimized to zero mean prediction error (PE) for each formula and lens–biometer combination. Primary outcome was mean absolute error (MAE), evaluated using heteroscedastic studentized bootstrap analysis (2000 iterations; Holm correction). Secondary outcomes included median absolute error, root-mean-square error, standard deviation of PE, and proportions of eyes within ±0.25diopter (D) and ±0.50D.

Results:

Included were 2550 patients (2550 eyes; mean age 74.0±10.0 years; 1068 male patients [41.9%]). Moussa produced a lower MAE (0.348 D) than Kane (0.359D), EVO (0.363D), PEARL-DGS (0.367D), K6 (0.367D), Hoffer-QST (0.377D), and BUII (0.379D; all p<.001). In short eyes (≤22.0mm), Moussa (0.412D) demonstrated lower error than K6 (p<.001), BUII (p<.001), Hoffer-QST (p=.008), EVO (p=.008), PEARL-DGS (p=.014), and Kane (p=.039). In long eyes (≥26.0 mm), Moussa (0.342 D) had lower error than Kane, BUII, Hoffer-QST (all p<.001), and K6 (p=.045), but not EVO (p=.068) or PEARL-DGS (p=.202).

Conclusions:

The Moussa formula demonstrated high predictive accuracy across multiple biometry platforms, with statistically lower overall MAE than the comparator formulas. Absolute differences from the closest-performing formulas were modest, and ±0.50D outcomes were similar to Kane and EVO. These findings support Moussa as a clinically relevant contemporary option for IOL power calculation, with encouraging results in short and long eyes requiring further targeted validation.