Comparing hemolysis among extracorporeal pumps: A large-scale retrospective study
Linda E. Sousse, Junya Hagiwara, Qianqian Liu, Timothy M. Maul, Joel E. Michalek, Jeffrey D. DellaVolpeBackground
Hemolysis is a frequent and clinically relevant complication during extracorporeal life support (ECLS, or ECMO), associated with acute kidney injury, systemic inflammation, and increased mortality. Direct comparisons between pumps regarding hemolysis are limited. This study evaluated and compared hemolysis profiles among several devices in clinical practice.
Methods
We conducted a retrospective, single-center cohort study adult patients who received veno-arterial (V-A) or veno-venous (V-V) ECMO between January 2020 and December 2023 (
Results
Partial thromboplastin time (
Conclusions
This study provides a clinical associations of the safety and performance of four centrifugal pumps in real-world ECMO practice in adults. We demonstrate that pump-specific RPM may be potentially correlated with hemolysis during ECMO in clinical practice, illustrating that the diagonal pump, which operated at the highest RPMs, was associated with the highest hemolysis in models without the RPM adjustment. Future studies are warranted to further clarify the impact of hemolysis on outcomes.