Comparative Performance of Oxygenators for Extracorporeal Membrane Oxygenation in Adults: A Retrospective Analysis
LINDA E. SOUSSE, JUNYA HAGIWARA, QIANQIAN LIU, TIMOTHY M. MAUL, JOEL E. MICHALEK, BRADFORD J. ANDERSON, JEFFREY D. DELLAVOLPE
Extracorporeal membrane oxygenation (ECMO) is a vital therapy for patients with severe cardiac and pulmonary failure. Despite technological advances, oxygenator dysfunction remains common, yet comparative clinical data across devices are limited. This study evaluated the performance of four contemporary adult oxygenators used in ECMO in a clinical, real-world setting. This retrospective single-center study included adult patients who received venovenous (VV) or venoarterial (VA) ECMO between January 2020 and December 2023 (N = 282). Oxygenators evaluated included the Medtronic Nautilus/Nautilus Smart, Getinge Heart Lung Support Module Advanced 7.0, Getinge Quadrox-iD Adult, and Fresenius X-Lung. Analyses were performed using propensity weighting and included adjustments for Acute Physiology and Chronic Health Evaluation II severity scores and baseline Charlson Comorbidity Index (CCI). No statistically significant differences in oxygen transfer across the membrane lung (VO
2
) were observed among oxygenator types after adjustment for covariates. VO
2
declined over time during support (