Balloon shaft fracture, deflation failure, and intracoronary entrapment: A stepwise percutaneous bailout – A case Report

Do Anh Nguyen, Tien Hoa Phan, Son Chau Do Truong

Abstract

Background

Balloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously.

Case summary

A 62-year-old man with multivessel coronary disease underwent staged PCI of an a left circumflex–obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully.

Discussion

Sequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.

More from our Archive