An analysis of initial force and moment delivery of different aligner materials
Thomas Wendl, Brigitte Wendl, Peter ProffAbstract
Objectives
The aim of this study was to clarify the applied initial forces and moments by different aligners of various materials and manufacturing methods.
Methods
The finite-element-method was used to analyze the forces and moments generated by the aligners on a maloccluded tooth. Plaster models of dental arches with a mesiorotated tooth 11 were fabricated, digitized and virtually analyzed. Four types of aligners with various layer thicknesses were selected: two splints with novel shape memory properties: a printable aligner made of the resin Tera Harz TC-85 DAC (Graphy Inc., South Korea) and a self-manufactured aligner consisting of the components polypropylene carbonate and thermoplastic polyurethane. The other two aligners were conventional, thermoformable aligners: CA® Pro Clear Aligner (Scheu Dental GmbH, Germany) and Erkodur-al (Erkodent Erich Kopp GmbH, Germany).
Results
The force and moment analyses showed that the thermoformable CA® Pro Clear Aligner exhibited the highest values. The thermoformed Erkodur-al aligner showed the lowest force loads for all layer thicknesses. The Graphy printed splint showed similar results compared to Erkodur-al at layer thicknesses of 0.40 mm and 0.50 mm.
Conclusions
To avoid periodontal overloading, aligners with lower force and moment delivery should be chosen for this type of tooth movement.