A Wearable Respiratory Monitor for Home-Based Screening and Stratification of Obstructive Sleep Apnea: A Pilot Study in Participants with Intermediate-to-High STOP-Bang Scores
Burcu Kolukisa Birgec, Beyza Toprak, Alexander Balfour MullenAccess to in-laboratory polysomnography (PSG) is restricted by prolonged waiting lists, and first-night effects can distort typical sleep architecture. This study evaluates the PneumoWave biosensor as a scalable, unobtrusive alternative for longitudinal single-channel home sleep apnea monitoring. Over three nights in an uncontrolled home environment, the biosensor’s respiratory rate agreement was evaluated against smartwatch-derived respiratory rate estimates, whilst apnea/hypopnea detection was compared with concurrent pulse oximetry. PneumoWave and smartwatch devices demonstrated good correlation (r = 0.870; p < 0.001). The PneumoWave device showed strong measurement agreement and provided a highly predictive screening pathway for patients with intermediate-to-high obstructive sleep apnea (OSA) risk. Longitudinal analysis confirmed consistent multi-night performance without first-night effect biases (ICC = 0.956, p < 0.001). Furthermore, its intuitive design yielded zero patient-induced setup errors, highlighting its operational robustness for self-administered use. Combining this continuous chest wall monitor with the STOP-Bang clinical questionnaire has the potential to provide an effective predictive screening pathway, improving community OSA screening and assisting clinical triage. Further validation against full polysomnography is warranted before clinical adoption.